Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Gould 220S: (1 year) CAL VER /5520 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 10 NUMBER OF LINES: 83 ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # Sub Gould 220S Full Scale Attenuator /5520 # Sub Gould 220S Linearity /5520 # Sub Gould 220S Sensitivity Ctrl Chk /5520 # Sub Gould 220S Square Wave Resp /5520 # Sub Gould 220S Frequency Resp /5520 # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 20 min. - REQUIRED 1.004 DISP Temperature: 18 - 28 degC 1.004 DISP Humidity: <75% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 RSLT =CHANNEL 1 TESTS 1.007 DISP Set CHANNEL 1 and CHANNEL 2 SENSITIVITY Controls to OFF. # Channel 1 1.008 MEM2 =Ch1 1.009 CALL Sub Gould 220S Full Scale Attenuator /5520 1.010 JMP 2.001 1.011 EVAL dummy 2.001 CALL Sub Gould 220S Sensitivity Ctrl Chk /5520 2.002 JMP 3.001 2.003 EVAL dummy 3.001 CALL Sub Gould 220S Linearity /5520 3.002 JMP 4.001 3.003 EVAL dummy 4.001 CALL Sub Gould 220S Square Wave Resp /5520 4.002 JMP 5.001 4.003 EVAL dummy 5.001 CALL Sub Gould 220S Frequency Resp /5520 5.002 JMP 6.001 5.003 EVAL dummy 6.001 DISP Set the CHART SPEED to STOP 6.002 RSLT = 6.003 DISP Set CHANNEL 1 and CHANNEL 2 SENSITIVITY Controls to OFF. # Channel 2 6.004 MEM2 =Ch2 6.005 CALL Sub Gould 220S Full Scale Attenuator /5520 6.006 JMP 7.001 6.007 EVAL dummy 7.001 CALL Sub Gould 220S Sensitivity Ctrl Chk /5520 7.002 JMP 8.001 7.003 EVAL dummy 8.001 CALL Sub Gould 220S Linearity /5520 8.002 JMP 9.001 8.003 EVAL dummy 9.001 CALL Sub Gould 220S Square Wave Resp /5520 9.002 JMP 10.001 9.003 EVAL dummy 10.001 CALL Sub Gould 220S Frequency Resp /5520 10.002 DISP SET the CHART SPEED to STOP. 10.003 END